SkyVision SDS System - Video Integration
Medical device recall Z-0109-2021. Published in FDA’s Enforcement Report on Oct 21, 2020.
No Watch here, and no alerts: openFDA asks that its enforcement data “should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall” (openFDA). For FDA’s own recall emails, subscribe at FDA. FDA’s own record is on accessdata.fda.gov.
The short answer
Published in FDA’s Enforcement Report on Oct 21, 2020. Copied from openFDA on Oct 20, 2026.
- Class II FDA’s class
- Terminated its status, in FDA’s file
- Nationwide FDA’s word, and 9 states named
- 36 units how much, in FDA’s words
Every line below is in FDA’s own words, with its source beside it. FDA’s terms are explained in FDA’s own words.
How serious is it, by FDA’s class?
Class II, the middle of FDA’s three classes.
“Products that might cause a temporary health problem, or pose only a slight threat of a serious nature. Example: a drug that is under-strength but that is not used to treat life-threatening situations.”
openFDA, in plain words, its enforcement report fields.
“Class II is a situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.”
FDA’s rule, 21 CFR 7.3(m)(2), eCFR.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
What’s wrong with it?
Intermittent failure of a signal being sent to a monitor from the touch panel display causing the displaying monitor to go Blank and the touch panel to display a Blank image when reviewing the target display. If this issue were to happen during a case, cause increase risk to patient due to hospital losing the visual image.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Which ones are recalled?
SkyVision SDS System - Video Integration Model: GC-SDS-BSCB-200 - Product Usage: is a high quality, self-contained video switching and control unit for use in the operating room.
36 units, in FDA’s words.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Codes to check against yours
Serial Numbers: DIPSG212460015 DIPSG212510043 DIPSG213160042 DIPSG214480117 DIPSG215040016 DIPSG219380004 DIPSG219380005 DIPSG219380006 DIPSG219380007 DIPSG219380008 DIPSG219380013 DIPSG219380014 DIPSG219380015 DIPSG219380016…
Every code shown here, 555 characters
Serial Numbers: DIPSG212460015 DIPSG212510043 DIPSG213160042 DIPSG214480117 DIPSG215040016 DIPSG219380004 DIPSG219380005 DIPSG219380006 DIPSG219380007 DIPSG219380008 DIPSG219380013 DIPSG219380014 DIPSG219380015 DIPSG219380016 DIPSG219380017 DIPSG219380018 DIPSG219380019 DIPSG219380020 DIPSG219380021 DIPSG219380022 DIPSG219380023 DIPSG219380024 DIPSG219380025 DIPSG219380026 DIPSG219420002 DIPSG219420003 DIPSG219420004 DIPSG219420005 DIPSG219420006 DIPSG219420007 DIPSG219470004 DIPSG219470008 DIPSG219470011 DIPSG219470013 DIPSG219470014 DIPSG219490014
Copied exactly as FDA lists them: lot and serial numbers, dates and other codes on the product or its label.
Where did it go?
FDA’s words for the “general area of initial distribution”: where the firm first shipped it.
US Nationwide distribution including in the states of CA, FL, IA, MI, MN, NY, PA, TX, WA.
The states and territories these words name
- California
- Florida
- Iowa
- Michigan
- Minnesota
- New York
- Pennsylvania
- Texas
- Washington
What FDA means by nationwide
“The term “nationwide” is defined to mean the fifty states or a significant portion. Note that subsequent distribution by the consignees to other parties may not be included.”
openFDA, its enforcement report fields.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
Is it still going on?
Terminated, in FDA’s file.
“FDA has determined that all reasonable efforts have been made to remove or correct the violative product in accordance with the recall strategy, and proper disposition has been made according to the degree of hazard.”
openFDA, on “Terminated”, its enforcement report fields.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
- The firm began the recall on Mar 10, 2020.
- FDA classified it on Oct 15, 2020.
- FDA’s Enforcement Report published it on Oct 21, 2020.
- FDA terminated it on Jun 1, 2021.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Who is recalling it?
- Skytron, Div. The KMW Group, Inc, of Grand Rapids, MI.
- Who began it, in FDA’s words: Voluntary: Firm initiated.
- How the firm first told its customers: Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
What this record can’t tell you
- Whether your own one is affected: compare the codes, dates and numbers above with yours. When unsure, ask the firm or the store.
- Where it went after the firm’s first shipments. FDA’s words cover the firm’s initial distribution, and, in FDA’s words, “subsequent distribution by the consignees to other parties may not be included.”
- How many have been returned, corrected or destroyed: FDA’s report doesn’t say.
- Whether anyone was hurt or fell ill: the report doesn’t say.
- How current it is: openFDA publishes FDA’s Enforcement Report weekly, after FDA does, and a firm can announce a recall weeks before it appears there. openFDA’s own caution: “This data should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall.”
- A product with no recall here isn’t safe by that: FDA’s report lists the recalls FDA classified, and when.
Hearth Record wrote these lines; the quotations are FDA’s.
Sources
-
U.S. Food and Drug Administration, Device recall enforcement reports.
Released Oct 14, 2026 and checked Oct 20, 2026. Updated weekly. Public domain (CC0 1.0, openFDA). FDA asks that nothing suggest it endorses this site.