Versys¿ Beaded Fullcoat Femoral Stem (POR FULL-CT FEM ST 11X160MM

Medical device recall Z-0070-2014. Published in FDA’s Enforcement Report on Nov 6, 2013.

FDA’s own record, on accessdata.fda.gov

No Watch here, and no alerts: openFDA asks that its enforcement data “should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall” (openFDA). For FDA’s own recall emails, subscribe at FDA. FDA’s own record is on accessdata.fda.gov.

The short answer

Published in FDA’s Enforcement Report on Nov 6, 2013. Copied from openFDA on Oct 20, 2026.

Every line below is in FDA’s own words, with its source beside it. FDA’s terms are explained in FDA’s own words.

How serious is it, by FDA’s class?

Class II, the middle of FDA’s three classes.

“Products that might cause a temporary health problem, or pose only a slight threat of a serious nature. Example: a drug that is under-strength but that is not used to treat life-threatening situations.”

openFDA, in plain words, its enforcement report fields.

“Class II is a situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.”

FDA’s rule, 21 CFR 7.3(m)(2), eCFR.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

What’s wrong with it?

The low density polyethylene (LDPE) bag used to package implants adheres to the highly polished surface of the devices.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Which ones are recalled?

Versys¿ Beaded Fullcoat Femoral Stem (POR FULL-CT FEM ST 11X160MM; VERSYS BEADED FC 11X200MM 8ST; POR FULLCT FEM ST 11 LOW EXT; POR FULLCT FEM ST 11 EXT; POR FLCT FEM STEM LM 11X160MM; POR FULLCT LM FEM ST 11 EXT POR FULL-CT FEM ST 12X160MM VERSYS BEADED FC 12X200MM 8 ST POR FULLCT FEM ST 12 LOW EXT POR FLCT FEM ST 12.0X190 STR POR FULLCT FEM ST 12 EXT POR FULL-CT FEM ST LM 12X160MM POR FULLCT LM FEM ST 12 EXT POR FULL-CT FEM ST 13X160MM VERSYS HIP SYSTEM REVISION FEMORAL BEADED FULLCOAT PLUS COLL POR FULLCT FEM ST 13 LOW EXT POR FLCT FEM ST 13.5X190 STR POR FULLCT FEM ST 13 EXT POR FULL-CT FEM ST LM 13X160MM POR FULLCT LM FEM ST 13 EXT POR FULL-CT FEM ST 14X160MM VERSYS BEADED FC 14X200MM 8 ST POR FULLCT FEM ST 14 LOW EXT POR FULLCT FEM ST 14 EXT POR FULL-CT FEM ST LM 14X160MM POR FULLCT LM FEM ST 14 EXT POR FULL-CT FEM ST 15X160MM VERSYS BEADED FC 15X200MM 8 ST POR FULLCT FEM ST 15 LOW EXT POR FLCT FEM ST 15.0X190 STR POR FLCT FEM ST 15.0X250 STR POR FULLCT FEM ST 15 EXT POR FULL-CT FEM ST LM 15X160MM POR FULLCT FEM ST 15 XEXT POR FULLCT LM FEM ST 15 EXT POR FULL-CT FEM ST 16X160MM VERSYS BEADED FC 16X200MM 8 ST POR FULLCT FEM ST 16 LOW EXT POR FLCT FEM ST 16.5X190 STR POR FLCT FEM ST 16.5X250 STR POR FULLCT FEM ST 16 EXT POR FULL-CT FEM ST LM 16X160MM POR FULLCT FEM ST 16 XEXT POR FULLCT LM FEM ST 16 EXT POR FULL-CT FEM ST 17X160MM VERSYS BEADED FC 17X200MM 8 ST POR FULLCT FEM ST 17 EXT POR FULL-CT FEM ST LM 17X160MM POR FULLCT FEM ST 17 XEXT POR FULLCT LM FEM ST 17 EXT POR FULL-CT FEM ST 18X160MM VERSYS BEADED FC 18X200MM 8 ST POR FLCT FEM ST 18.0X190 STR POR FLCT FEM ST 18.0X250 STR POR FULLCT FEM ST 18 EXT POR FULL-CT FEM ST LM 18X160MM POR FULLCT FEM ST 18 XEXT POR FULLCT LM FEM ST 18 EXT POR FULL-CT FEM ST 19X160MM POR FLCT FEM ST 19.5X190 STR POR FLCT FEM ST 19.5X250 STR POR FULLCT FEM ST 19 EXT POR FLCT FEM STEM LM 19X160MM POR FULLCT FEM ST 19 XEXT POR FULLCT LM FEM ST 19 EXT POR FULL-CT FEM ST 20X160MM VERSYS BEADED FC 20X200MM 8 ST POR FULLCT FEM ST 20 EXT POR FLCT FEM STEM LM 20X160MM POR FULLCT FEM ST 20 XEXT POR FULLCT LM FEM ST 20 EXT POR FULL-CT FEM ST 21X160MM POR FLCT FEM ST 21.0X190 STR POR FLCT FEM ST 21.0X250 STR POR FULLCT FEM ST 21 EXT POR FLCT FEM STEM LM 21X160MM POR FULLCT FEM ST 21 XEXT POR FULLCT LM FEM ST 21 EXT POR FULLCT FEM STEM 22X160MM POR FLCT FEM ST 22.5X190 STR POR FLCT FEM ST 22.5X250 STR POR FULLCT FEM ST 22 EXT POR FLCT FEM STEM LM 22X160MM POR FULLCT FEM ST 22 XEXT POR FULLCT LM FEM ST 22 EXT. Used in total hip replacement.

6380, in FDA’s words.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Codes to check against yours

Part Number Range: 00-7843-011/022-06/56

Source:

Copied exactly as FDA lists them: lot and serial numbers, dates and other codes on the product or its label.

Where did it go?

FDA’s words for the “general area of initial distribution”: where the firm first shipped it.

Worldwide Distribution: US (nationwide) and countries of: Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Dominican Republic, Ecuador, El Salvador, Germany, Guatemala, Honduras, Hong Kong, India, Israel, Jamaica, Japan, Korea, Malaysia, Mexico, New Zealand, Nicaragua, Panama, Peru, Singapore, Taiwan, Thailand, and Venezula.

These words name no state.

What FDA means by nationwide

“The term “nationwide” is defined to mean the fifty states or a significant portion. Note that subsequent distribution by the consignees to other parties may not be included.”

openFDA, its enforcement report fields.

The words say it went outside the United States too.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

Is it still going on?

Terminated, in FDA’s file.

“FDA has determined that all reasonable efforts have been made to remove or correct the violative product in accordance with the recall strategy, and proper disposition has been made according to the degree of hazard.”

openFDA, on “Terminated”, its enforcement report fields.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

  • The firm began the recall on Aug 30, 2013.
  • FDA classified it on Oct 25, 2013.
  • FDA’s Enforcement Report published it on Nov 6, 2013.
  • FDA terminated it on Sep 25, 2014.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Who is recalling it?

  • Zimmer, Inc., of Warsaw, IN.
  • Who began it, in FDA’s words: Voluntary: Firm initiated.
  • How the firm first told its customers: Letter.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

What this record can’t tell you

  • Whether your own one is affected: compare the codes, dates and numbers above with yours. When unsure, ask the firm or the store.
  • Where it went after the firm’s first shipments. FDA’s words cover the firm’s initial distribution, and, in FDA’s words, “subsequent distribution by the consignees to other parties may not be included.”
  • How many have been returned, corrected or destroyed: FDA’s report doesn’t say.
  • Whether anyone was hurt or fell ill: the report doesn’t say.
  • How current it is: openFDA publishes FDA’s Enforcement Report weekly, after FDA does, and a firm can announce a recall weeks before it appears there. openFDA’s own caution: “This data should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall.”
  • A product with no recall here isn’t safe by that: FDA’s report lists the recalls FDA classified, and when.
Source:

Hearth Record wrote these lines; the quotations are FDA’s.

Sources

  • U.S. Food and Drug Administration, Device recall enforcement reports.

    Released Oct 14, 2026 and checked Oct 20, 2026. Updated weekly. Public domain (CC0 1.0, openFDA). FDA asks that nothing suggest it endorses this site.

Source:

Recalls of food, medicines, medical devices, pet food and animal drugs are from the U.S. Food and Drug Administration (FDA). FDA does not endorse Hearth Record. openFDA: “Do not rely on openFDA to make decisions regarding medical care.” Copied on Oct 20, 2026.

How to read a record

Four things every record does, shown on an example record made up for this guide.

An example record, made up for this guide

An example nursing home

Made up for this guide: no real place is shown

Not fully inspected in over two years (CMS flag)

11 health citations at the latest inspection cycle

The state average is 14.3.

Source:

CMS, State and US Averages and Health Deficiencies. Released Oct 28, 2026. Updated monthly.

From the homeNone yet. Replies and paid listings aren’t open, and when they open they will sit here, below the record.

1 of 4

Every fact names its source.

Who published it, when, and how often it changes sit beside each fact, with a link to the agency’s own file.

Sign in

A free account keeps your watch list, shortlists, comparisons and saved searches on every device, and can email you when something you watch changes. Browsing never needs one.