Hearth Record

Monarch Inflation Device MAP 112 Access-9 Large Bore Hemostasis Value Metal Insertion Tool Torque DeviceREF…

Medical device recall Z-0058-2025. Published in FDA’s Enforcement Report on Oct 23, 2024.

FDA’s own record, on accessdata.fda.gov

No Watch here, and no alerts: openFDA asks that its enforcement data “should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall” (openFDA). For FDA’s own recall emails, subscribe at FDA. FDA’s own record is on accessdata.fda.gov.

The short answer

Published in FDA’s Enforcement Report on Oct 23, 2024. Copied from openFDA on Oct 20, 2026.

Every line below is in FDA’s own words, with its source beside it. FDA’s terms are explained in FDA’s own words.

How serious is it, by FDA’s class?

Class II, the middle of FDA’s three classes.

“Products that might cause a temporary health problem, or pose only a slight threat of a serious nature. Example: a drug that is under-strength but that is not used to treat life-threatening situations.”

openFDA, in plain words, its enforcement report fields.

“Class II is a situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.”

FDA’s rule, 21 CFR 7.3(m)(2), eCFR.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

What’s wrong with it?

Due to a manufacturing process resulting in holes/breaches in the Tyvek sterile barrier.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Which ones are recalled?

Monarch Inflation Device MAP 112 Access-9 Large Bore Hemostasis Value Metal Insertion Tool Torque DeviceREF IN2230 Injector And Syringe, Angiographic/Syringe, Balloon Inflation - The 30 ATM Monarch syringe is used to inflate and deflate balloon angioplasty catheters and to measure the pressure within the balloon during the procedure. It is also used to inject fluids into the body and to monitor the pressure of that fluid.

41,411 devices, in FDA’s words.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Codes to check against yours

Catalog Number: IN2230/B UDI-DI code: 00884450298039 Lot Number: H2287420 H2292343 H2298614 H2305033 H2309342 H2312490 H2320981 H2326309 H2331538 H2337132 H2343845 H2348707 H2353507 H2360151 H2379871 H2385818 H2395792 H2398304 H2413090…

Every code shown here, 802 characters

Catalog Number: IN2230/B UDI-DI code: 00884450298039 Lot Number: H2287420 H2292343 H2298614 H2305033 H2309342 H2312490 H2320981 H2326309 H2331538 H2337132 H2343845 H2348707 H2353507 H2360151 H2379871 H2385818 H2395792 H2398304 H2413090 H2416387 H2421986 H2432426 H2435878 H2442961 H2446109 H2453675 H2457933 H2458274 H2466447 H2474177 H2478393 H2483314 H2486956 H2492840 H2497955 H2503844 H2507119 H2511930 H2517517 H2522877 H2528486 H2540011 H2545410 H2550516 H2554634 H2560206 H2564557 H2568221 H2574048 H2579553 H2584200 H2589842 H2594940 H2599814 H2604238 H2610524 H2614052 H2618903 H2622808 H2629605 H2633298 H2639295 H2645646 H2649893 H2655505 H2660061 H2664662 H2669779 H2673553 H2675990 H2683251 H2701290 H2709357 H2713928 H2733074 H2739475 H2746858 H2751186 H2760546 H2765638 H2789805 H2819396

Source:

Copied exactly as FDA lists them: lot and serial numbers, dates and other codes on the product or its label.

Where did it go?

FDA’s words for the “general area of initial distribution”: where the firm first shipped it.

Worldwide - U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries of Belgium, Canada, France, Japan, Monaco, Netherlands, Qatar, United Arab Emirates, United Kingdom.

The states and territories these words name

46 states and territories.

Every one of them
  • Alabama
  • Arkansas
  • Arizona
  • California
  • Colorado
  • Connecticut
  • Florida
  • Georgia
  • Iowa
  • Idaho
  • Illinois
  • Indiana
  • Kansas
  • Kentucky
  • Louisiana
  • Massachusetts
  • Maryland
  • Michigan
  • Minnesota
  • Missouri
  • Mississippi
  • Montana
  • North Carolina
  • North Dakota
  • Nebraska
  • New Hampshire
  • New Jersey
  • New Mexico
  • Nevada
  • New York
  • Ohio
  • Oklahoma
  • Oregon
  • Pennsylvania
  • Rhode Island
  • South Carolina
  • South Dakota
  • Tennessee
  • Texas
  • Utah
  • Virginia
  • Vermont
  • Washington
  • Wisconsin
  • West Virginia
  • Wyoming

What FDA means by nationwide

“The term “nationwide” is defined to mean the fifty states or a significant portion. Note that subsequent distribution by the consignees to other parties may not be included.”

openFDA, its enforcement report fields.

The words say it went outside the United States too.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

Is it still going on?

Ongoing, in FDA’s file.

“A recall which is currently in progress.”

openFDA, on “On-Going”, its enforcement report fields.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

  • The firm began the recall on Aug 30, 2024.
  • FDA classified it on Oct 11, 2024.
  • FDA’s Enforcement Report published it on Oct 23, 2024.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Who is recalling it?

  • Merit Medical Systems, Inc., of South Jordan, UT.
  • Who began it, in FDA’s words: Voluntary: Firm initiated.
  • How the firm first told its customers: Letter.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

What this record can’t tell you

  • Whether your own one is affected: compare the codes, dates and numbers above with yours. When unsure, ask the firm or the store.
  • Where it went after the firm’s first shipments. FDA’s words cover the firm’s initial distribution, and, in FDA’s words, “subsequent distribution by the consignees to other parties may not be included.”
  • How many have been returned, corrected or destroyed: FDA’s report doesn’t say.
  • Whether anyone was hurt or fell ill: the report doesn’t say.
  • How current it is: openFDA publishes FDA’s Enforcement Report weekly, after FDA does, and a firm can announce a recall weeks before it appears there. openFDA’s own caution: “This data should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall.”
  • A product with no recall here isn’t safe by that: FDA’s report lists the recalls FDA classified, and when.
Source:

Hearth Record wrote these lines; the quotations are FDA’s.

Sources

  • U.S. Food and Drug Administration, Device recall enforcement reports.

    Released Oct 14, 2026 and checked Oct 20, 2026. Updated weekly. Public domain (CC0 1.0, openFDA). FDA asks that nothing suggest it endorses this site.

Source:

Recalls of food, medicines, medical devices, pet food and animal drugs are from the U.S. Food and Drug Administration (FDA). FDA does not endorse Hearth Record. openFDA: “Do not rely on openFDA to make decisions regarding medical care.” Copied on Oct 20, 2026.

How to read a record

Four things every record does, shown on an example record made up for this guide.

An example record, made up for this guide

An example nursing home

Made up for this guide: no real place is shown

Not fully inspected in over two years (CMS flag)

11 health citations at the latest inspection cycle

The state average is 14.3.

Source:

CMS, State and US Averages and Health Deficiencies. Released Oct 28, 2026. Updated monthly.

From the homeNone yet. Replies and paid listings aren’t open, and when they open they will sit here, below the record.

1 of 4

Every fact names its source.

Who published it, when, and how often it changes sit beside each fact, with a link to the agency’s own file.

Sign in

A free account keeps your watch list, shortlists, comparisons and saved searches on every device, and can email you when something you watch changes. Browsing never needs one.