STERRAD 200 Sterilizer (Single Door)

Medical device recall Z-0043-2014. Published in FDA’s Enforcement Report on Oct 30, 2013.

FDA’s own record, on accessdata.fda.gov

No Watch here, and no alerts: openFDA asks that its enforcement data “should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall” (openFDA). For FDA’s own recall emails, subscribe at FDA. FDA’s own record is on accessdata.fda.gov.

The short answer

Published in FDA’s Enforcement Report on Oct 30, 2013. Copied from openFDA on Oct 20, 2026.

Every line below is in FDA’s own words, with its source beside it. FDA’s terms are explained in FDA’s own words.

How serious is it, by FDA’s class?

Class II, the middle of FDA’s three classes.

“Products that might cause a temporary health problem, or pose only a slight threat of a serious nature. Example: a drug that is under-strength but that is not used to treat life-threatening situations.”

openFDA, in plain words, its enforcement report fields.

“Class II is a situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.”

FDA’s rule, 21 CFR 7.3(m)(2), eCFR.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

What’s wrong with it?

Advanced Sterilization Products (ASP) is recalling the STERRAD 200 Sterilization System because it may emit an odor or smell into the surrounding environment that is not typical to normal operating conditions.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Which ones are recalled?

STERRAD 200 Sterilizer (Single Door), Part Number: 10201. STERRAD 200 Sterilizer (Double Door), Part Number: 10202. The STERRAD Sterilization System is a low temperature general purpose sterilizer used to sterilize heat and moisture sensitive reusable medical devices.

1,305 units total (654 units in US , 651 OUS ), in FDA’s words.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Codes to check against yours

Lot A: 0009*ENG 0003*FG 0048*FG 0050*MRB 0237*YAMATO 0239*YAMATO 0097*FSRIP 0097*FG 315*TTL 316*TTL 317*TTL 318*TTL 314*FG 325*TTL 326*TTL 334*FG 335*FG 329*TTL 345*TTL 357*FG 359*FG 358*FG 344*TTL 351*FG 347*TTL 352*TTL 348*TTL 348*FG…

Every code shown here, 1,996 characters

Lot A: 0009*ENG 0003*FG 0048*FG 0050*MRB 0237*YAMATO 0239*YAMATO 0097*FSRIP 0097*FG 315*TTL 316*TTL 317*TTL 318*TTL 314*FG 325*TTL 326*TTL 334*FG 335*FG 329*TTL 345*TTL 357*FG 359*FG 358*FG 344*TTL 351*FG 347*TTL 352*TTL 348*TTL 348*FG 360*TTL 361*TTL 364*FG 362*TTL 383*YAMATO 365*TTL 368*TTL 371*FG 369*TTL 370*TTL 374*TTL 378*TTL 379*TTL 397*TTL 376*TTL 398*TTL 400*TTL 401*TTL 402*TTL 386*TTL 403*TTL 404*TTL 406*TTL 405*TTL 399*TTL 419*TTL 388*TTL 420*TTL 421*TTL 0239*CD33 422*TTL 423*TTL 429*YAMATO 424*TTL 439*YAMATO 425*TTL 427*TTL 346*TTL 423*TTL 372*TTL 430*TTL 428*TTL 395*TTL 431*TTL 433*TTL 447*YAMATO 432*TTL 434*TTL 435*TTL 437*TTL 436*TTL 453*YAMATO 454*YAMATO 452*YAMATO 440*TTL 412*TTL 455*YAMATO 456*YAMATO 468*FG 460*FG 458*FG 459*FG 462*FG 463*FG 464*FG 473*FG 470*FG 465*FG 488*YAMATO 476*FG 481*FG 487*YAMATO 502*FG 482*FG 501*FG 484*FG 485*FG 486*FG 483*FG 479*TTL 493*FG 492*FG 495*YAMATO 494*FG 509*YAMATO 504*TTL 503*TTL 505*TTL 506*TTL 507*TTL 512*TTL 524*TTL 540*YAMATO 538*YAMATO 539*YAMATO 537*YAMATO 541*YAMATO 508*TTL 520*TTL 519*TTL 518*TTL 516*TTL 521*TTL 522*TTL 523*TTL 556*TTL 525*TTL 526*TTL 524*TTL 528*TTL 529*TTL 530*TTL 532*TTL 531*TTL 515*FG 533*TTL 534*TTL 535*TTL 557*TTL 536*TTL 568*YAMATO 570*YAMATO 575*YAMATO 543*TTL 545*TTL 544*TTL 571*YAMATO 546*TTL 548*TTL 547*TTL 549*TTL 550*TTL 554*TTL 553*TTL 552*TTL 558*TTL 555*TTL 581*YAMATO 559*TTL 560*TTL 561*TTL 585*YAMATO 562*TTL 563*TTL 576*TTL 579*TTL 593*TTL 595*TTL 591*YAMATO 590*YAMATO 596*FG 597*TTL 606*YAMATO 602*TTL 601*TTL 599*TTL 598*TTL 607*YAMATO 604*TTL 603*TTL 612*FG 616*YAMATO 611*TTL 628*TTL 615*TTL 629*TTL 632*TTL 630*TTL 631*TTL 605*TTL 625*YAMATO 621*YAMATO 643*TTL 633*TTL 644*TTL 645*TTL 656*TTL 652*TTL 655*TTL 658*TTL 646*TTL 654*FG 659*FG 660*TTL 662*TTL 680*TTL 663*TTL 665*TTL 666*TTL 664*TTL 667*TTL 683*TTL 685*TTL 686*TTL 684*TTL 650*YAMATO 620*YAMATO 653*TTL 679*TTL 682*TTL 690*TTL 692*TTL 691*TTL 693*TTL 694*TTL 695*TTL 696*TTL 689*TTL 688*TTL 687*TTL 661*TTL…

FDA’s list runs to 8,560 characters. The first part is shown; the rest is on FDA’s record.

Source:

Copied exactly as FDA lists them: lot and serial numbers, dates and other codes on the product or its label.

Where did it go?

FDA’s words for the “general area of initial distribution”: where the firm first shipped it.

Worldwide Distribution - USA (nationwide) and the countries of EA (Europe Middle East and Asia), APAC (Asia Pacific), Japan and Canada

These words name no state.

What FDA means by nationwide

“The term “nationwide” is defined to mean the fifty states or a significant portion. Note that subsequent distribution by the consignees to other parties may not be included.”

openFDA, its enforcement report fields.

The words say it went outside the United States too.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

Is it still going on?

Terminated, in FDA’s file.

“FDA has determined that all reasonable efforts have been made to remove or correct the violative product in accordance with the recall strategy, and proper disposition has been made according to the degree of hazard.”

openFDA, on “Terminated”, its enforcement report fields.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

  • The firm began the recall on Sep 18, 2013.
  • FDA classified it on Oct 22, 2013.
  • FDA’s Enforcement Report published it on Oct 30, 2013.
  • FDA terminated it on Apr 11, 2014.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Who is recalling it?

  • Advanced Sterilization Products, of Irvine, CA.
  • Who began it, in FDA’s words: Voluntary: Firm initiated.
  • How the firm first told its customers: Letter.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

What this record can’t tell you

  • Whether your own one is affected: compare the codes, dates and numbers above with yours. When unsure, ask the firm or the store.
  • Where it went after the firm’s first shipments. FDA’s words cover the firm’s initial distribution, and, in FDA’s words, “subsequent distribution by the consignees to other parties may not be included.”
  • How many have been returned, corrected or destroyed: FDA’s report doesn’t say.
  • Whether anyone was hurt or fell ill: the report doesn’t say.
  • How current it is: openFDA publishes FDA’s Enforcement Report weekly, after FDA does, and a firm can announce a recall weeks before it appears there. openFDA’s own caution: “This data should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall.”
  • FDA’s list of codes runs to 8,560 characters; the first part is shown here, and the rest is on FDA’s record.
  • A product with no recall here isn’t safe by that: FDA’s report lists the recalls FDA classified, and when.
Source:

Hearth Record wrote these lines; the quotations are FDA’s.

Sources

  • U.S. Food and Drug Administration, Device recall enforcement reports.

    Released Oct 14, 2026 and checked Oct 20, 2026. Updated weekly. Public domain (CC0 1.0, openFDA). FDA asks that nothing suggest it endorses this site.

Source:

Recalls of food, medicines, medical devices, pet food and animal drugs are from the U.S. Food and Drug Administration (FDA). FDA does not endorse Hearth Record. openFDA: “Do not rely on openFDA to make decisions regarding medical care.” Copied on Oct 20, 2026.

How to read a record

Four things every record does, shown on an example record made up for this guide.

An example record, made up for this guide

An example nursing home

Made up for this guide: no real place is shown

Not fully inspected in over two years (CMS flag)

11 health citations at the latest inspection cycle

The state average is 14.3.

Source:

CMS, State and US Averages and Health Deficiencies. Released Oct 28, 2026. Updated monthly.

From the homeNone yet. Replies and paid listings aren’t open, and when they open they will sit here, below the record.

1 of 4

Every fact names its source.

Who published it, when, and how often it changes sit beside each fact, with a link to the agency’s own file.

Sign in

A free account keeps your watch list, shortlists, comparisons and saved searches on every device, and can email you when something you watch changes. Browsing never needs one.