GE Healthcare, Precision 500D, Legacy, RFX, and SFX X
Medical device recall Z-0012-2014. Published in FDA’s Enforcement Report on Oct 23, 2013.
No Watch here, and no alerts: openFDA asks that its enforcement data “should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall” (openFDA). For FDA’s own recall emails, subscribe at FDA. FDA’s own record is on accessdata.fda.gov.
The short answer
Published in FDA’s Enforcement Report on Oct 23, 2013. Copied from openFDA on Oct 20, 2026.
- Class II FDA’s class
- Terminated its status, in FDA’s file
- District of Columbia the one state its words name
- 810 (783 US, 27 OUS) how much, in FDA’s words
Every line below is in FDA’s own words, with its source beside it. FDA’s terms are explained in FDA’s own words.
How serious is it, by FDA’s class?
Class II, the middle of FDA’s three classes.
“Products that might cause a temporary health problem, or pose only a slight threat of a serious nature. Example: a drug that is under-strength but that is not used to treat life-threatening situations.”
openFDA, in plain words, its enforcement report fields.
“Class II is a situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.”
FDA’s rule, 21 CFR 7.3(m)(2), eCFR.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
What’s wrong with it?
GE Healthcare has recently become aware of a potential safety issue involving the spotfilm device power assist handle. There is a potential for the internal bolts on the power assist handle to be loose or to loosen and/or fall out. This issue, if present, can result in a loss of calibration and offset of the power assist handle and/or, under extreme condition, sticking of the handle in the ON p
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Which ones are recalled?
GE Healthcare, Precision 500D, Legacy, RFX, and SFX X. This table is intended for use in a diagnostic X-ray system to support patients during general purpose radiological procedures in the horizontal, vertical and trendelenburg positions. - The Expedio 500D is designed to perform radiographic and fluoroscopic x-ray examinations. - The Precision 500D R&F X-ray System is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic examinations, including general R&F and pediatric examinations. The Precision 500D R&F X-ray Systems are intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic and fluoroscopic exposures of the whole body, skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position.
810 (783 US, 27 OUS), in FDA’s words.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Codes to check against yours
Serial # System ID 00000976624WK7 202884RF2410 00000977748WK3 202884RF2411 00000955594WK7 205877LRF2 00000965983WK0 207373RF1 00000969026WK4 215748LEG1 00000960066WK9 219416ADV 00000975683WK4 228809SR2 00000966338WK6 256235ORF1…
Every code shown here, 1,994 characters
Serial # System ID 00000976624WK7 202884RF2410 00000977748WK3 202884RF2411 00000955594WK7 205877LRF2 00000965983WK0 207373RF1 00000969026WK4 215748LEG1 00000960066WK9 219416ADV 00000975683WK4 228809SR2 00000966338WK6 256235ORF1 00000967770WK9 301896RF 00000961102WK1 302654RF 00000972383WK4 303440BCHRM4 00000964398WK2 303730R3 00000958345WK1 304256RX1A 00000978812WK6 361552LEGD1 00000970377WK8 404321LEG2 00000973016WK9 408259RF1 00000963305WK8 408977RF5 00000963524WK4 414266RM03 00000963118WK5 417533BR2 00000969537WK0 507372LEG 00000959491WK2 518382SRM2 00000976668WK4 540862ARF6 00000966773WK4 561468RM4 00000972380WK0 606285LEGACY 00000966622WK3 619421RF1 00000957841WK0 708216GIRF1 00000983335WK1 713527PRF1 00000964482WK4 775738RM2 00000971676WK2 804289HDF4 00000978181WK6 804747DHP1 00000994573WK4 815740RF1 00000961772WK1 817472LEG1 00000963293WK6 850747ER1 00000975653WK7 856778RF2 00000965747WK9 864295GRRM4 00000973832WK9 904508LEGACY 00000962333WK1 904819RF1 00000970009WK7 907276RM3 00000953103WK9 912633MR3 00000978182WK4 915685ADV3 00000978314WK3 916784LEG 00000977043WK9 936321LEG1 00000963738WK0 941798RM2 00000982588WK6 956388RF1 00000966348WK5 972566RM11 00000962460WK2 RADNETRF130 00000963491WK6 RADNETRF137 00000978497WK6 651232RF4 00000983536WK4 703689RM03 00000011284GP7 210297BMCRF11 00000011281GP3 210297SEBRF3 00000985785WK5 281348P5001 00000014340GP4 308398RF 00000014337GP0 412641RF1 00000984897WK9 435637RM4 00000013713GP3 561863RF1 00000013860GP2 661633XR2 00000014347GP9 713636LBJ7 00000988380WK2 717272RF2 00000011676GP4 720854RF1 00000012513GP8 727823RM3 00000013045GP0 770793P500D 00000012418GP0 781744RF3 00000012521GP1 816348BRF1 00000009196GP7 817255NHPD1 00000013285GP2 828298RF4 00000011737GP4 850325RF4 00000013257GP1 850494WF4 00000014350GP3 856325P500 00000989504WK6 913676SRM4 00000013271GP2 956350VRF3 00000990179WK4 PC7479XR03 00000012826GP4 00105FLU01 00001009797WK0 303825P500 00000009180M31 202745PRE2 00000022650M36 207498P500 00000001835M38…
FDA’s list runs to 20,967 characters. The first part is shown; the rest is on FDA’s record.
Copied exactly as FDA lists them: lot and serial numbers, dates and other codes on the product or its label.
Where did it go?
FDA’s words for the “general area of initial distribution”: where the firm first shipped it.
Worldwide Distribution - USA (nationwided) including DC., and the countries of AUSTRALIA, CANADA, CHILE, GERMANY, CHANA, GUATEMALA, ICELAND, ISRAEL, ITALY, JAMAICA, JAPAN, RUSIAN FEDERATION, UNITED ARAB EMIRATES and UNITED KINGDOM.
The states and territories these words name
- District of Columbia
The words say it went outside the United States too.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Is it still going on?
Terminated, in FDA’s file.
“FDA has determined that all reasonable efforts have been made to remove or correct the violative product in accordance with the recall strategy, and proper disposition has been made according to the degree of hazard.”
openFDA, on “Terminated”, its enforcement report fields.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
- The firm began the recall on Aug 30, 2013.
- FDA classified it on Oct 17, 2013.
- FDA’s Enforcement Report published it on Oct 23, 2013.
- FDA terminated it on May 22, 2014.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Who is recalling it?
- GE Healthcare, LLC, of Waukesha, WI.
- Who began it, in FDA’s words: Voluntary: Firm initiated.
- How the firm first told its customers: Letter.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
What this record can’t tell you
- Whether your own one is affected: compare the codes, dates and numbers above with yours. When unsure, ask the firm or the store.
- Where it went after the firm’s first shipments. FDA’s words cover the firm’s initial distribution, and, in FDA’s words, “subsequent distribution by the consignees to other parties may not be included.”
- How many have been returned, corrected or destroyed: FDA’s report doesn’t say.
- Whether anyone was hurt or fell ill: the report doesn’t say.
- How current it is: openFDA publishes FDA’s Enforcement Report weekly, after FDA does, and a firm can announce a recall weeks before it appears there. openFDA’s own caution: “This data should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall.”
- FDA’s list of codes runs to 20,967 characters; the first part is shown here, and the rest is on FDA’s record.
- A product with no recall here isn’t safe by that: FDA’s report lists the recalls FDA classified, and when.
Hearth Record wrote these lines; the quotations are FDA’s.