Renal Care Group-Southeast Tyler

A dialysis facility in Tyler, Texas 75703

3826 Troup Highway, Suite # L, Tyler, Texas 75703. (903) 581-5252.

CMS’s own record, on Care Compare

Watch and the shortlist are kept with a free account, on every device. Your watch list. CMS’s own record is on Care Compare.

The short answer

Record last released by CMS on Oct 28, 2026.

The last 24 months 12 deficiencies in 3 months of the last 36. Last cited May 2025. Most serious: Standard-level. Every month

Each figure’s source sits beside it, in its section below.

What these words mean3
Star ratings
CMS rates some kinds of place from 1 to 5 stars, more stars meaning better on CMS’s own measures for that kind: for a nursing home, its health inspections, staffing and quality measures. A rating is CMS’s summary of those measures; it doesn’t promise good care. Source: CMS, Care Compare and the Five-Star Quality Rating System.
Inspection (survey)
A visit by state inspectors, often unannounced, who check a place against the rules it must meet. For nursing homes and other places Medicare pays, CMS calls these visits surveys: a standard survey of a nursing home comes at least every 15 months, and a complaint survey whenever a complaint calls for one. Source: CMS; 42 CFR 488.308.
Deficiency, citation, violation
What inspectors write up when a place didn’t meet a rule: the rule, what they saw, and when. CMS calls it a deficiency; states may say citation or violation. It is the inspectors’ finding, not a court’s, and the place is usually asked for a plan to correct it. Source: CMS, statement of deficiencies (Form CMS-2567); each state’s licensing agency.

What did inspectors find?

The last inspection on record was on May 5, 2025: a complaint inspection.

12 deficiencies cited in the last three years.

Source:

CMS, QCOR Dialysis Facility (ESRD) 2567 statements of deficiencies. Released Oct 28, 2026. Updated monthly.

When those deficiencies happened

12 deficiencies in 3 months of the last 36. Last cited May 2025. Most serious: Standard-level. The last 3 years, to Oct 30, 2026.

12 deficiencies by month over the last 3 years. Each cell is one month.
JanFebMarAprMayJunJulAugSepOctNovDec
2023 NovDec
2024 JanFebMarAprMayJunJulAugSepOctNovDec
2025 JanFebMarAprMayJunJulAugSepOctNovDec
2026 JanFebMarAprMayJunJulAugSepOct

Hover or focus a month to read it.

No visit on record Visited, nothing cited A stronger orange is more serious, on its source’s own levels, read onto four bands: from the least serious it publishes to the most
Source:

Hearth Record drew this from the dates and levels in the source’s own records.

Each deficiency, as cited

Rule titles as CMS’s inspectors cite them, set in sentence case.

Source:

CMS, QCOR Dialysis Facility (ESRD) 2567 statements of deficiencies. Released Oct 28, 2026. Updated monthly.

May 5, 2025: complaint inspection5 deficiencies
  • Ic - Sanitary Environment

    The inspectors’ words, 591 words

    “IC-SANITARY ENVIRONMENT CFR(s): 494.30 The dialysis facility must provide and monitor a sanitary environment to minimize the transmission of infectious agents within and between the unit and any adjacent hospital or other public areas. This Standard is not met as evidenced by: 48653 Based on observation, record review, and staff interview, the facility staff failed to ensure a clean and sanitary environment was maintained to reduce the transmission of infectious disease in 1 (in-center treatment area) of 1 clinical area. The facility staff failed to follow the “General Cleanliness and Infection Control Guidelines” policy. During a tour of the facility’s in-center treatment area on 05/05/2025 at 10:30 AM, the surveyor observed the following infection control issues, There was a mix of sterile patient supplies and dirty items stored together on top of the counters in the in-center treatment area in front of dialysis treatment stations #17-22. There was a staff members personal lotion bottled stored on the counter directly next to sterile patient supplies. There was a large blood spill on the floor in front of station # 3. The blood was on the dialysis machine and splattered on the floor around the dialysis machine. There was a patient at the station and the station next the spill at the time of the observation. There was a build-up of dust and dirt particles on the counter tops located in front of station # 2. There was a stethoscope laying on the sink labeled “clean”. It is unknown if the stethoscope had been cleaned and disinfected. There was a large pile of blue and white dialysis clamps laying on the counter. It was unknown if the dialysis clamps were clean or dirty because they were not labeled and stored appropriately. There was a build-up of dust and dirt on top of the emergency code cart. The cart had a large plastic covering the top of the cart but the plastic had been removed and not placed back on the cart appropriately which allowed dust to collect on the cart where patient supplies are stored. There was a pair of black scissors laying in the area labeled “clean sink”. The scissors had a build-up of orange and brown substance on them. The counters had a build-up of supplies and equipment which caused unnecessary clutter. The counters could not be adequately cleaned and disinfected due to the clutter. A review of the facilities policy, “General Cleanliness and Infection Control Guidelines” with a date of 05/05/2025 revealed, “Purpose: The purpose of this policy is to provide guidance for FKC staff on preventing the spread of infectious disease and maintaining a clean, safe, and aesthetically pleasant environment for patients, staff, and visitors .... Policy: All areas must be kept clean and organized, including but not limited to treatment areas, water/supply rooms, offices, and restrooms. Walkways must be kept clear of debris and free of clutter. Facilities must adhere to any state=specific regulations that pertain to facility cleaning and disinfection. Eating, drinking, applying cosmetics or lip balm, and handling contact lenses are prohibited in work areas where there is a reasonable likelihood of occupational exposure ...Facility staff are accountable for cleaning rooms/areas not assigned to contracted cleaning staff. Such cleaning should be done regularly using a schedule developed by the facility ....” An interview was conducted with Staff # 5 on 05/05/2025 at 11:30 AM. Staff # 5 confirmed the infection control findings inside the in-center treatment area. Staff # 5 confirmed that the facility staff failed to follow the “General Cleanliness and Infection Control Guidelines” policy.”

    Rule V0111. Standard-level. From a complaint.

  • Ic - Clean/Dirty Areas, Med Prep Area, No Common Med Carts

    The inspectors’ words, 431 words

    “IC-CLEAN/DIRTY;MED PREP AREA;NO COMMON CARTS CFR(s): 494.30(a)(1)(i) Clean areas should be clearly designated for the preparation, handling and storage of medications and unused supplies and equipment. Clean areas should be clearly separated from contaminated areas where used supplies and equipment are handled. Do not handle and store medications or clean supplies in the same or an adjacent area to that where used equipment or blood samples are handled. When multiple dose medication vials are used (including vials containing diluents), prepare individual patient doses in a clean (centralized) area away from dialysis stations and deliver separately to each patient. Do not carry multiple dose medication vials from station to station. Do not use common medication carts to deliver medications to patients. If trays are used to deliver medications to individual patients, they must be cleaned between patients. This Standard is not met as evidenced by: 48653 Based on observation, record review, and staff interview, the facility staff failed to ensure that medications maintained in the facility were properly safeguarded and not accessible to unauthorized persons. The facility staff failed to follow the “Medication Policy”. During a tour of the facility’s in-center treatment area on 05/05/2025, the surveyor observed the emergency code cart unlocked with the second drawer open. The second drawer contained emergency medications that were not secured. The medications were accessible to unauthorized persons. The surveyor observed a patient # 2’s prescription for Velphoro 500 mg (milligram) chewable tablets on the nurses’ station. There were no staff members in the vicinity of the medication, and the medication was accessible to unauthorized persons. A review of the facility policy titled “MEDICATION POLICY” revealed, “PURPOSE: To provide guidance for medication management in the facility and to provide guidance for the safe and aseptic preparation of all medications. POLICY: 1. The Administrator/designee is responsible for supervising the handling, storing, disposing, administering, and controlling of medications and performs a monthly audit and inventory ... 11. All refrigerated medications, for example, erythropoietin stimulating agents (ESA) and vaccines are to be locked at the close of each business day or if not under supervision by the licensed teammate or per state regulations. Non-refrigerated medications are to be stored in cabinet(s) and locked at the close of each business day or if not under supervision by the licensed teammate or per state regulations. 23. Medications and needle/syringe storage are done in accordance with state regulations.” An interview was conducted with Staff # 5 on 05/05/2025 at 11:30 AM. Staff # 5 confirmed that the medications in the in-center treatment area should be secure and safeguarded from unauthorized persons.”

    Rule V0117. Standard-level. From a complaint.

  • Mixing Systems - Labeling

    The inspectors’ words, 640 words

    “MIXING SYSTEMS-LABELING CFR(s): 494.40(a) 5.4.4.1 Mixing systems: labeling Labeling strategies should permit positive identification by anyone using the contents of mixing tanks, bulk storage/dispensing tanks, and small containers intended for use with a single hemodialysis machine. Mixing tanks: Prior to batch preparation, a label should be affixed to the mixing tank that includes the date of preparation and the chemical composition or formulation of the concentrate being prepared. This labeling should remain on the mixing tank until the tank has been emptied. Bulk storage/dispensing tanks: These tanks should be permanently labeled to identify the chemical composition or formulation of their contents. Concentrate jugs: At a minimum, concentrate jugs should be labeled with sufficient information to differentiate the contents from other concentrate formulations used at the facility. This Standard is not met as evidenced by: 48653 Based on observation, record review, and staff interview, the facility staff failed to ensure 3 of 3 acid concentrate jugs were properly labeled and stored in a secure environment. The facility staff failed to follow the “Acid Concentrate Storage and Handling” policy. During a tour of the facility’s in-center treatment area on 05/05/2025 at 10:30 AM, the surveyor observed three 1-gallon acid concentrate jugs stored in the middle of the in-center treatment area on a blue cart with wheels. The acid concentrates were all opened with the seal broken and stored on the same shelf, despite being different concentrations. Jug # 1 was labeled 3.0 K/2.5 Ca and was dated 04/23/2025. There was no time on the jug as required in the facility’s policy. Jug # 2 was labeled 2.0 K/3.0 Ca and was dated 05/01/2025. There was no time on the jug as required in the facility’s policy. Jug # 3 was labeled 3.0 K/3.0 Ca and was dated 03/13/2025. There was no time on the jug as required in the facility’s policy. A review of the facility’s policy, “Acid Concentrate Storage and Handling” with a date of 09/03/2024 revealed, “Policy: Concentrates must be stored in a secured area accessible only to authorized staff members ... The concentrate containers must be kept in a climate-controlled area within the temperature ranges described on the container’s labeling. Store concentrate formulas by type, formulation, and outer case identifiers, such as formula number. Container labels must be clearly visible, be in good condition, and readable. For incenter facilities ...label shelving with formula type. Label all acid containers as outlined in the Concentrate Labeling Requirements policy. Discontinued or expired concentrate formulas must be removed from inventory. Unused acid concentrate that cannot be transferred or returned must be disposed of in accordance with state and local regulations. Blending of unlike formulas is strictly prohibited .... Labeling will occur when: ... 1-gallon: if unused acid will be stored for further use, record the date and time it was opened and the initials of the staff member that opened the container... Post treatment: If acid concentrate will be stored for future use: Externally disinfect the gallon jug with wand still in place using 1:100 bleach cloth or hospital-approved disinfectant. Keep the wand in place until the wand can be removed and replaced with a clean cap. Return the container to the designated storage area. Open, labeled concentrates must be stored separately on a properly labeled location. Do not return the opened acid to the same shelf as the unopened concentrate. Remove the wand assembly, replace the clean cap on the container, rinse wand with RO water, hang vertically, and allow to dry. Once the wand is dry, store in a covered container labeled “clean wands”. If not able to meet requirements, acid concentrate must be discarded after treatment ...” An interview was conducted with Staff # 5 on 05/05/2025 at 11:30 AM, which confirmed the facility staff members were not properly labeling and storing the acid concentrates according to the facility’s policy.”

    Rule V0228. Standard-level. From a complaint.

  • Mixing Systems - Cleaning

    The inspectors’ words, 286 words

    “MIXING SYSTEMS-CLEANING CFR(s): 494.40(a) 6.4.1 Mixing systems: cleaning Concentrate mixing equipment should be either: (1) completely emptied, cleaned, and disinfected according to the manufacturer’s instructions; or (2) cleaned and disinfected using a procedure demonstrated by the facility to be effective in routinely producing concentrate meeting [these regulations related to allowable bacterial and endotoxin levels]. The disinfection data should be recorded for each ...disinfection cycle using a dedicated log. This Standard is not met as evidenced by: 48653 Based on observation, record review, and staff interview, the facility staff failed to ensure the acid pick-up wands were properly disinfected and stored. The facility staff failed to follow the “Acid Concentrate Storage and Handling” policy. During a tour of the facility’s in-center treatment area on 05/05/2025 at 10:30 AM, the surveyor observed there were several acid pick-up wands lying out on the counter and next to the sink labeled “dirty sink”. It was unknown if the acid pick-up wands had been properly cleaned and disinfected because they were not labeled and were not stored vertically for proper drying. A review of the facility’s policy, “Acid Concentrate Storage and Handling” with a date of 09/03/2024 revealed, “Policy: Concentrates must be stored in a secured area accessible only to authorized staff members ... Post Treatment: If concentrate will be stored for future use: ... Remove the wand assembly, replace the clean cap on the container, rinse wand with RO water, hang vertically and allow to dry. Once wand is dry, store in a covered container labeled, “Clean Wands” .... An interview was conducted with Staff # 5 on 05/05/2025 at 11:30 AM which confirmed the facility staff were not handling the acid pick up wands according to the facility’s policy.”

    Rule V0230. Standard-level. From a complaint.

  • Dialysate Proportioning - Monitor Ph/Conductivity

    The inspectors’ words, 334 words

    “DIALYS PROPORT-MONITOR PH/CONDUCTIVITY CFR(s): 494.40(a) 5.6 Dialysate proportioning: monitor pH/conductivity It is necessary for the operator to follow the manufacturer’s instructions regarding dialysate conductivity and to measure approximate pH with an independent method before starting the treatment of the next patient. This Standard is not met as evidenced by: 48653 Based on observation, record review, and staff interview, the facility staff failed to follow the manufacturer’s instructions for use (IFU) for the conductivity testing meter. During a tour of the facility’s in-center treatment area on 05/05/2025 at 10:30 AM, the surveyor observed 4 plastic spray bottles labeled as “RO water”. The bottles did not have the current date written on them to indicate that the RO (reverse osmosis) water had been changed. RO water expires after 24 hours and needs to be changed daily. The bottles had the following dates, Bottle # 1 “04/12/2025” Bottle # 2 “04/29/2025” Bottle # 3 “04/29/2025” Bottle # 4 “03/14/2025” A review of the sticker located on the side of the spray bottles revealed, “Frequent rinsing is the most effective means of preventing hard deposits and bacteria from collecting on conductivity cell electrodes and altering the readings. Rinse the instrument after use. Empty this bottle DAILY and refill it with fresh RO water.” A review of the facility’s policy, “Myron L D6 Measuring Procedures” with a revision date of 11/04/2024 revealed, “Measuring Dialysate Conductivity and pH: ...3. Rinse sample cell according to dialysate meter IFU ...” A review of the “Digital Dialysate Meter Operations Manual” revealed, “Care and maintenance: Dialysate meters should be rinsed with clean water after use ...” An interview was conducted with Staff # 2 on 05/05/2025 at 11:45 AM. Staff # 5 was asked if the conductivity meters are rinsed with the RO water in the spray bottles that were outdated. Staff # 2 stated, “We no longer have to rinse the conductivity meters”. Staff # 2 confirmed that the facility staff were not following the facility’s policy or the manufacturer’s instructions for use.”

    Rule V0250. Standard-level. From a complaint.

Jan 13, 2025: complaint inspection6 deficiencies
  • Ic - Sanitary Environment

    The inspectors’ words, 481 words

    “IC-SANITARY ENVIRONMENT CFR(s): 494.30 The dialysis facility must provide and monitor a sanitary environment to minimize the transmission of infectious agents within and between the unit and any adjacent hospital or other public areas. This Standard is not met as evidenced by: 48653 Based on observation, record review, and staff interview, the facility staff failed to ensure a clean and sanitary environment was monitored and maintained in 1 (in-center treatment area) of 1 clinical area. Also, the facility staff failed to follow the facility policy, “General Cleanliness and Infection Control Guidelines”. This deficient practice had the likelihood to result in the transmission of infectious agents within the facility and between patients, staff members, and members of the public. The CDC defines a “sanitary environment” as an environment that meets the “Standard Precautions” for an inpatient hospital setting plus the more stringent precautions that are recommended for hemodialysis units because of the increased potential for contamination with blood and pathogenic microorganisms. During a tour of the facility’s In-Center Treatment Area on 01/13/2025 at 09:45 AM with Staff # 5, the surveyor observed the following infection control issues: * The patient chair located at Station # 8 had a large rip in the seat. There was no way to ensure the chair was adequately cleaned and disinfected due to the damage to the chair’s material. * A patient’s neck pillow was sitting on the counter of the nurse’s station directly next to the “clean” dialysis clamps. * There was an opened AV (arteriovenous) fistula needle sitting on the counter. It was unclear if the needle had been used on a patient or when the needle was opened. *The counters located next to the nurse’s station were cluttered with patient equipment and supplies. There was dust built up on the counter. *There were multiple blood pressure cuffs with unknown substances on the Velcro portion of the cuff. A review of the facility’s policy, “General Cleanliness and Infection Control Guidelines” with a published date of 11/04/2024 revealed, “PURPOSE: The purpose of this policy is to provide guidance for FKC staff on preventing the spread of infectious disease and maintaining a clean, safe, and aesthetically pleasant environment for patients, staff, and visitors ... POLICY: *All areas must be kept clean and organized, including but not limited to treatment areas, water/supply rooms and offices. Walkways must be kept clear of debris and free of clutter. *Facilities must adhere to any state-specific regulations that pertain to facility cleaning and disinfection ... *Facility staff are accountable for cleaning rooms/areas not assigned to contracted cleaning staff. Such cleaning should be done regularly using a schedule developed by the facility. *Supplies or patient belongings should not be kept or stored behind the machine at the patient station ....” An interview with Staff # 5 on 01/13/2025 at 10:30 AM confirmed the infection control issues found in the facility’s in-center treatment area.”

    Rule V0111. Standard-level. From a complaint.

  • Ic - Wear Gloves/Hand Hygiene

    The inspectors’ words, 425 words

    “IC-WEAR GLOVES/HAND HYGIENE CFR(s): 494.30(a)(1) Wear disposable gloves when caring for the patient or touching the patient’s equipment at the dialysis station. Staff must remove gloves and wash hands between each patient or station. This Standard is not met as evidenced by: 48653 Based on observation, policy review, and staff interview, 1 (Staff # 24) of 2 direct care staff members failed to change gloves and perform hand hygiene between patient encounters and before touching dialysis equipment and sterile supplies. Also, the facility’s direct care staff failed to follow the facility’s “Cleaning and Disinfecting the Dialysis Station” policy. Findings include: During a tour of the facility’s in-center treatment area on 01/13/2025 at 10:00 AM with Staff # 5, the surveyor observed Staff # 24 (CCHT) providing care to the patient at dialysis station #27. Staff # 24 was observed touching the dialysis machine and keyboard with the same gloves that were used to touch the patient and the patient’s dialysis chair. Staff # 24 did not perform hand hygiene or change gloves between patient contact and contact with the dialysis machine and chair. Also, Staff # 24 used the same gloves to provide care to the patient at dialysis station # 28. Staff # 24 never performed hand hygiene or changed gloves between the two patient interactions. A review of the facility’s policy, “Cleaning and Disinfecting the Dialysis Station” with a published date of 09/05/2023 revealed, “PURPOSE: The purpose of this policy is to provide guidelines to prevent the spread of infectious disease in accordance with appropriate regulations, and to maintain a clean, safe, and aesthetically pleasant environment for patients, staff, and visitors. BACKGROUND: “The Centers for Medicare and Medicaid Services (CMS) has regulations that in order to prevent cross-contamination, a dialysis station must be cleaned and disinfected between dialysis patients. The chair and dialysis equipment are used by multiple patients during a treatment day and it is critical that these items be thoroughly cleaned and disinfected between uses. Also, the nature of dialysis treatments with frequent exposure to blood and body fluids, close proximity of patients and staff, and the immunocompromised status of dialysis patients make dialysis a high-risk area for spreading infectious disease ... Hand hygiene must be performed after contact with the electronic devices and before contact with the patient ...” An interview with Staff # 5 on 01/13/2025 at 10:45 AM confirmed Staff # 24 did not change gloves and complete hand hygiene at the appropriate times and intervals as required in the facility policy, “Cleaning and Disinfecting the Dialysis Station”.”

    Rule V0113. Standard-level. From a complaint.

  • Ic - Clean/Dirty Areas, Med Prep Area, No Common Med Carts

    The inspectors’ words, 463 words

    “IC-CLEAN/DIRTY;MED PREP AREA;NO COMMON CARTS CFR(s): 494.30(a)(1)(i) Clean areas should be clearly designated for the preparation, handling and storage of medications and unused supplies and equipment. Clean areas should be clearly separated from contaminated areas where used supplies and equipment are handled. Do not handle and store medications or clean supplies in the same or an adjacent area to that where used equipment or blood samples are handled. When multiple dose medication vials are used (including vials containing diluents), prepare individual patient doses in a clean (centralized) area away from dialysis stations and deliver separately to each patient. Do not carry multiple dose medication vials from station to station. Do not use common medication carts to deliver medications to patients. If trays are used to deliver medications to individual patients, they must be cleaned between patients. This Standard is not met as evidenced by: 48653 Based on observation, record review, and staff interview, the facility staff failed to ensure that 4 of 4 medications maintained in the facility were properly safeguarded and not accessible to unauthorized persons. The facility staff failed to follow the “Medication Preparation and Administration” policy. Findings include: During a tour of the facility’s in-center treatment area on 01/13/2025 at 09:45 AM with Staff # 5, the surveyor observed four pre-made treatment initiation packs (packs of supplies used to initiate dialysis treatment) sitting on the counter next to the main nurse’s station. The pre-made packs contained syringes of pre-drawn Heparin (a medication used to thin blood). The medication was not labeled with the date and time it was prepared. There was no way to determine how long the Heparin had been sitting on the counter or when it was prepared. The Heparin was not secured and was available for unauthorized persons to access. A review of the facility’s policy, “Medication Preparation and Administration” with a published date of 02/06/2023 revealed, “PURPOSE: To administer medications with the goals of staff and patient safety, optimal therapeutic response, and infection control ... PRE-DRAWING MEDICATIONS: Medications may be pre-drawn up to 4 hours prior to administration. These pre-drawn medications shall be labeled and must be kept under the preparer’s control or in a locked designated medication storage area or refrigerated, if necessary, until delivery to the appropriate patient for administration ... Expiration dates for all stored medications are to be monitored on a monthly basis. Expired medications are to be discarded via Fresenius Kidney Care off-site return program or in accordance with local and/or state laws ...” An interview with Staff # 5 on 01/13/2025 at 09:55 AM confirmed that the facility staff failed to ensure medications were locked up and not accessible to unauthorized persons. Also, Staff # 5 confirmed the syringes were not properly labeled with the date and time of preparation.”

    Rule V0117. Standard-level. From a complaint.

  • Ic - Transducer Protectors: Not Wetted/Changed

    The inspectors’ words, 508 words

    “IC-TRANSDUCER PROTECTORS-NOT WETTED/CHANGED CFR(s): 494.30(a)(1)(i) Use external venous and arterial pressure transducer filters/protectors for each patient treatment to prevent blood contamination of the dialysis machines’ pressure monitors. If the external transducer protector becomes wet, replace immediately and inspect the protector. If fluid is visible on the side of the transducer protector that faces the machine, have qualified personnel open the machine after the treatment is completed and check for contamination. This includes inspection for possible blood contamination of the internal pressure tubing set and pressure sensing port. If contamination has occurred, the machine must be taken out of service and disinfected using either 1:100 dilution of bleach (300-600 mg/L free chlorine) or a commercially available, EPA-registered tuberculocidal germicide before reuse. Change filters/protectors between each patient treatment, and do not reuse them. Internal transducer filters do not need to be changed routinely between patients. This Standard is not met as evidenced by: 48653 Based on observation, policy review, and staff interviews, the facility staff failed to identify 2 (Station # 5 and # 12) of 2 wet (wet with patient blood or other fluid) external transducers (pressure monitors) located on the dialysis machines and failed to follow the facility policy, “Maintenance of Transducer Protectors Procedure.” This deficient practice could cause harm to all patients receiving dialysis at the facility by introducing patient blood into the dialysis machine. This could cause contamination of the dialysis machine and allow other patients to become exposed to blood contamination. Findings include: During a tour of the facility’s In-Center Treatment Area on 01/13/2025 at 10:00 AM, the surveyor observed the dialysis machines at stations # 5 and # 12 have a wet external transducer. The external transducer was wet with the patient’s blood due to high venous pressures during treatment. The surveyor alerted Staff # 5 to the wet transducers. A review of the facility’s policy, “Maintenance of Transducer Protectors Procedures” with a published date of 11/07/2022 revealed, “PURPOSE: The purpose of this policy is to ensure accurate measurement of arterial and venous pressures and prevent cross-contamination during hemodialysis by maintaining clean, dry transducer protectors. BACKGROUND: The transducer protectors must be dry to provide accurate venous and arterial pressures. Cross-contamination can occur if the internal transducer becomes contaminated with blood and is not disinfected or changed before the machine is used on another patient. POLICY: The transducer protector must be changed immediately when contaminated with saline or blood. If the back of the transducer protector is contaminated with blood, the internal transducer must be checked by biomedical staff prior to use of the machine for another patient for contamination and changed or disinfected as necessary. The internal transducer will be examined on at least an annual basis in the home, semiannual basis, or 2000 hours, whichever comes first in the clinic, and any times maintenance is performed by the biomedical staff as part of routine maintenance ...” An interview with Staff # 5 on 01/13/2025 at 10:00 AM confirmed the transducers at stations # 5 and # 12 were wet with the patient’s blood.”

    Rule V0120. Standard-level. From a complaint.

  • Mixing Systems - Labeling

    The inspectors’ words, 505 words

    “MIXING SYSTEMS-LABELING CFR(s): 494.40(a) 5.4.4.1 Mixing systems: labeling Labeling strategies should permit positive identification by anyone using the contents of mixing tanks, bulk storage/dispensing tanks, and small containers intended for use with a single hemodialysis machine. Mixing tanks: Prior to batch preparation, a label should be affixed to the mixing tank that includes the date of preparation and the chemical composition or formulation of the concentrate being prepared. This labeling should remain on the mixing tank until the tank has been emptied. Bulk storage/dispensing tanks: These tanks should be permanently labeled to identify the chemical composition or formulation of their contents. Concentrate jugs: At a minimum, concentrate jugs should be labeled with sufficient information to differentiate the contents from other concentrate formulations used at the facility. This Standard is not met as evidenced by: 48653 Based on observation, record review, and staff interview, the facility staff failed to ensure 4 of 4 open acid concentrate jugs were labeled with the date of opening. The facility staff failed to follow the policy, “Acid Concentrate Storage and Handling”. Findings include: During a tour of the facility’s in-center treatment area on 01/13/2025 at 09:45 AM with Staff # 5, the surveyor observed 4 acid-contrate jugs on a black cart. The seal on the jugs had been broken. The surveyor was unable to determine when the jugs had been opened or what the solution in the jug was because there was no label, date, or time. A review of the facility’s policy titled, “Acid Concentrate Storage and Handling” revealed the following: “The purpose of this policy is to provide direction for concentrate storage and handling. Store concentrate formulas by type, formulas and outer case identifiers, such as formula number. Outer box or container labels must be clearly visible, shall be in good condition and readable. Storage areas must use appropriate GHS labeling as stated in OSHA Hazard Communication Program. Label all acid containers as outlined in the Concentrate Labeling Requirements policy. Discontinued or expired concentrate formulas must be removed from inventory. Unused acid concentrate that cannot be transferred or returned, must be disposed of in accordance with state and local regulations. Concentrates must be stored in a secured area accessible only to authorized staff members or, if in the home setting, patient and/or care partner. Every 30 days, the reusable concentrate container must be cleaned: Discard any remaining concentrate. Thoroughly rinse inside and outside of container with reverse osmosis (RO) or deionized (DI) water. Inspect for foreign materials. Place on a shelf inverted to allow to dry Each container will be assigned and labeled with a permanent identification number to facilitate the tracking of containers. Containers will be stored capped to prevent accumulation of dust, dirt or other contaminants. Containers, wands and caps that are stored will be rinsed, allowed to air dry and stored in a manner that will prevent accumulation of dust, dirt or other contaminants.” An interview with Staff # 5 on 01/13/2025 at 10:30 AM confirmed the acid jugs were opened without the appropriate labeling.”

    Rule V0228. Standard-level. From a complaint.

  • Pe - Hemodialysis Patients In View During Treatments

    The inspectors’ words, 302 words

    “PE-HD PTS IN VIEW DURING TREATMENTS CFR(s): 494.60(c)(4) Patients must be in view of staff during hemodialysis treatment to ensure patient safety, (video surveillance will not meet this requirement). This Standard is not met as evidenced by: 48653 Based on observation, record review, and staff interview, the facility staff failed to ensure 2 (Patient # 3 and Patient # 6) of 2 patient’s arteriovenous (AV) access was visible throughout the entire dialysis treatment. Ensuring proper visualization of the arteriovenous fistula access is important for the early recognition and prevention of bleeding from the patient’s access site. Findings include: During a tour of the facility’s in-center treatment area on 01/13/2025 at 9:30 AM with Staff # 5, the surveyor observed the following: Patient # 3 receiving dialysis treatment at Station # 1. The patient’s arteriovenous access was covered with a personal blanket. There was no documentation in the patient’s medical record to support the patient’s education on the importance of always keeping the access visible. The surveyor observed the patient’s access covered for more than 5 minutes. Patient # 6 was receiving dialysis treatment at Station # 20. The patient’s arteriovenous access was covered with a personal blanket. A review of the facility’s policy, “Monitoring During Patient Treatment” revealed, “PURPOSE: The purpose of this policy is to provide direction for monitoring dialysis patients during treatment. POLICY: Monitor the patient at the initiation of treatment and every 30 minutes, or more frequently as necessary ... Access: Observe and document at the initiation of dialysis and at every safety check that all connections are secure and visible ...Ensure access remains uncovered throughout the treatment ...” An interview with Staff # 5 on 01/23/2025 at 10:00 AM confirmed that the facility staff failed to ensure that all patient’s access was visible throughout the entire dialysis treatment.”

    Rule V0407. Standard-level. From a complaint.

Sep 23, 2024: complaint inspection1 deficiency
  • Rn, Msw, & Rd Available To Meet Patient Needs

    The inspectors’ words, 283 words

    “GOV-RN, MSW, & RD AVAIL TO MEET PT NEEDS CFR(s): 494.180(b)(1) The governing body or designated person responsible must ensure that- The registered nurse, social worker and dietitian members of the interdisciplinary team are available to meet patient clinical needs; This Standard is not met as evidenced by: 48653 Based on staff interviews, the facility’s leadership failed to ensure the social worker had a maximum patient load of 125 patients and available clerical support to meet every patient’s psychological and social needs. Findings include: The facility treated patients Monday through Saturday and had a total patient census of 89. During an interview with Staff # 4 (Licensed Social Worker) on 09/23/2024 at 11:00 AM, Staff # 4 reported, in addition to his full-time commitment at the facility, he was also providing crisis staffing support to 2 other dialysis facilities, located 50 miles away. Staff # 4 reported the combined patient load he was supporting at the other 2 facilities was 70 patients. Staff # 4 acknowledged he was over the maximum patient load of 125 patients. Staff # 4 stated, “We are short 4 social workers in our region”. Staff # 4 (Licensed Social Worker) had a total patient load of 159 patients which exceeds the State maximum of 125 patients. Staff # 4 was unavailable for an in-person interview at the facility because he was working at one of the other dialysis facilities. Staff # 4 does not work at the facility on Mondays and is not present to provide support to patients on Mondays. During an interview with Staff # 1 (Clinical Manager) on 09/23/2024 at 11:15 AM, Staff # 1 acknowledged the Social Worker was above the maximum patient load.”

    Rule V0758. Standard-level. From a complaint.

All 3 inspections in the last three years
Inspections on record
DateVisitWhat the record says
May 5, 2025Complaint inspection5 deficiencies cited
Jan 13, 2025Complaint inspection6 deficiencies cited
Sep 23, 2024Complaint inspection1 deficiency cited

Is it certified?

  • Medicare-certified dialysis facility, certified or last recertified on Apr 23, 2004.
  • Run for profit.
  • No deficiency on record in the last three years.
  • Chain: Fresenius Medical Care.
Source:

CMS, Dialysis Facility - Listing by Facility. Released Jul 15, 2026. Updated quarterly.

CMS, QCOR Dialysis Facility (ESRD) 2567 statements of deficiencies. Released Oct 28, 2026. Updated monthly.

Has it been penalized?

  • CMS doesn’t publish fines or other penalties for dialysis facilities.
  • 12 deficiencies cited in the last three years came from inspections that included a complaint investigation.
Source:

CMS, QCOR Dialysis Facility (ESRD) 2567 statements of deficiencies. Released Oct 28, 2026. Updated monthly.

What does it offer?

  • Offers in-center hemodialysis.
  • Doesn’t offer peritoneal dialysis, home hemodialysis training and a shift starting at 5 p.m. or later.
  • 24 dialysis stations.
  • CMS doesn’t publish staffing levels for dialysis facilities.
  • In CMS’s survey, patients rate its staff 1 of 5.
Source:

CMS, Dialysis Facility - Listing by Facility and Patient survey (ICH CAHPS) - Facility. Released Jul 15, 2026. Updated quarterly.

Who owns it?

  • Part of the Fresenius Medical Care chain, as CMS lists it.
  • CMS publishes no owners file for dialysis facilities, so no other owner is named.
Source:

CMS, Dialysis Facility - Listing by Facility. Released Jul 15, 2026. Updated quarterly.

How do patients fare?

  • CMS rates it 4 of 5 stars (Jan 1, 2021 to Dec 31, 2024).
  • Compared by CMS with what it expects for the facility’s patients: patient survival, as expected; hospital admissions, as expected; readmissions to hospital, as expected; blood transfusions, as expected; emergency department visits, as expected; emergency department visits within 30 days of a hospital stay, as expected; bloodstream infections, as expected; patients dialysing through a fistula, as expected; patients put on a kidney transplant waitlist in their first year of dialysis, as expected; patients on a kidney transplant waitlist, as expected.
  • Patients rate it 2 of 5 in CMS’s survey of in-center hemodialysis patients (43 interviews, Oct 18, 2024 to Jul 11, 2025): 47% rated the facility 9 or 10 out of 10.
Source:

CMS, Dialysis Facility - Listing by Facility and Patient survey (ICH CAHPS) - Facility. Released Jul 15, 2026. Updated quarterly.

Ratings and measures, as published
MeasureRating
Star ratingCMS’s star rating for dialysis facilities, 1 to 54 of 5
Patient survey ratingCMS star rating from its survey of the facility’s in-center hemodialysis patients (ICH CAHPS), 1 to 52 of 5

How does it compare nearby?

  • Of the 6 dialysis facilities CMS lists in Smith County, 6 have a star rating from CMS: 1 rated 5 of 5, 2 rated 4 of 5, 3 rated 3 of 5.
  • CMS rates this one 4 of 5.
Source:

CMS, Dialysis Facility - Listing by Facility. Released Jul 15, 2026. Updated quarterly.

What the record doesn’t say

  • CMS doesn’t publish fines or other penalties for dialysis facilities, so none are shown.
  • Only complaints that led to a citation appear; complaints that didn’t aren’t published.
  • QCOR says it has no dialysis inspection reports after February 19, 2026, while it moves to a new system: a later inspection is not shown yet.
  • CMS doesn’t publish staffing levels for dialysis facilities.
  • CMS publishes no owners file for dialysis facilities: the chain it lists is the only owner named.

Sources

If you believe something here is wrong, a correction starts with the agency whose record this copies: CMS. CMS’s own record is on Care Compare. If Hearth Record copied something wrongly, it will fix it and log the correction publicly, with its date: how corrections work.

Source:

This copy of the record was made on Oct 30, 2026.

From the facility

The public record above comes first. Anything the facility adds will appear here, below it, and say who wrote it.

Replies

None. Replies aren’t open yet. When they open, the facility will be able to answer any finding for free, and the reply will show beside that finding, unedited and dated.

Paid listing

None. Paid listings aren’t open yet. When they open, anything paid for will say “Paid by the facility”, and paying will never change a finding, a rating or the order of results.

How to read a record

Four things every record does, shown on an example record made up for this guide.

An example record, made up for this guide

An example nursing home

Made up for this guide: no real place is shown

Not fully inspected in over two years (CMS flag)

11 health citations at the latest inspection cycle

The state average is 14.3.

Source:

CMS, State and US Averages and Health Deficiencies. Released Oct 8, 2026. Updated monthly.